Remote Heart Rate Sensors Can be Biased Against Darker Skin. A UCLA Team Offers a Solution

By combining two technologies, camera and radar, researchers boosted accuracy across a diverse range of skin tones
Heart Rate Sensor
 Visual Machines Group/UCLA

The mobile multimodal sensing platform uses camera and radar to remotely collect plethysmography data

Aug 24, 2022

Originally posted on UCLA Newsroom

As telemedicine has grown more popular, so have devices that allow people to measure their vital signs from home and transmit the results by computer to their doctors. Yet in many cases, obtaining accurate remote readings for people of color has proved a persistent challenge.

Take remote heart rate measurements, for example, which rely on a camera sensing subtle changes in the color of a patient’s face caused by fluctuations in the flow of blood beneath their skin. These devices, part of an emerging class of remote technologies, consistently have trouble reading color changes in people with darker skin tones, said Achuta Kadambi, an assistant professor of electrical and computer engineering at the UCLA Samueli School of Engineering.

Kadambi and his team have now developed a remote diagnostic technique that overcomes this implicit bias against darker skin while also making heart rate readings more accurate for patients across the full range of skin tones. Their secret? Combining the light-based measurements of a camera with radio-based measurements from radar.

The researchers presented their findings, recently published in the journal ACM Transactions on Graphics, at the SIGGRAPH 2022 conference in Vancouver, British Columbia. The conference, held both virtually and in-person, is organized annually by members of the Association for Computing Machinery.

The advance could lead to new classes of high-performing medical devices and remote technologies that are more accurate and equitable, the researchers said, allowing doctors and health care systems to remotely monitor patients with confidence, both in clinical settings and from patients’ homes.

“In the larger picture, this work shows that practical and innovative engineering solutions can address persistent biases in medical devices,” Achuta Kadambi said

“In the larger picture, this work shows that practical and innovative engineering solutions can address persistent biases in medical devices,” said Kadambi, who is also a member of the California NanoSystems Institute at UCLA. “But that first requires an acknowledgement that such bias means the current best technology may not be the best for everyone. Through thoughtful design, we can find equitable solutions that perform as well or better.”

The UCLA team’s fusion of two techniques shows a promising path toward achieving those goals, said Kadambi, who is also an assistant professor of computer science and principal investigator on the research. As head of the Visual Machines Group at UCLA, he has written about different types of biases in medical devices and how to fix them.

In developing their new technology, the researchers first showed that the remote sensing device itself was the source of the bias, demonstrating in their paper that higher levels of melanin, natural pigments in the skin, interfere with what is known as the photoplethysmography, or PPG, signal that is used in current camera-based remote heart rate measurements.

PPG signaling is also used to measure heart rate through devices like pulse oximeters, which clamp onto a patient’s finger, as well as some wearable commercial products and smartwatch-powered apps. These devices emit light onto the skin and sense changes in the amount of light reflected back by circulating blood just below the surface. That reflected light produces the PPG signal, a measure of a patient’s heart rate.

Heart Rate Sensor
From left: UCLA researchers Pradyumna Chari, Alexander “Sasha” Vilesov, Adnan Armouti, Dr. Laleh Jalilian and Achuta Kadambi with the new device

Previous efforts to address skin tone biases in such technologies have generally looked to correct them through additional programming or by expanding baseline standards using a more diverse range of skin tones. But neither of these approaches target the real issue, Kadambi said, which is the physics of the device itself.

The UCLA researchers instead turned to another technology that can give an estimate of heart rate: radar. At 77 gigahertz, radar can sense subtle changes in the displacement of the chest from heartbeats. And while this method overcomes the issue of skin tone bias, it is less reliable than PPG signaling. However, they found success by combining these two different modes of sensing — camera and radar — and refining them through machine learning to work in concert.